Product details
Veterinary Syringe Pump Features
Veterinary infusion pump is with the classic vertical design. The software compensation error is within±5% after calibration. Adjustable level setting and occlusion pressure release function is make it commonly acceptable by the veterinary hospital. Double CPU and numeric keypad make it safe and more user friendly.
Adjustable Levels Setting | 10 levels setting for occlusion and air bubble volume |
Strong Battery Backup | 3000mAh battery and backup time more than 10hrs. |
User Friendly | Big Screen and numeric keypad |
OOcclusion Release Function | Occlusion pressure release and restarting function |
Infusion Mode | Volume mode, time mode |
Specification of Veterinary Syringe Pump
Item | Specifification |
Injection Mode | Volume Mode, Time Mode, Drug Mode |
IV Sets | Standard single use IV sets (10,15,20,60d/mL, with diameter 3.4 ~ 4.5mm) |
Flow Rate | 1.0ml/h~1200ml/h |
Increment | 1.0-99.9ml/h in 0.1ml/h step 100-1200ml/ in 1ml/h step |
Deviation | Flow Rate:≤5%(after calibration) |
VTBI | 0.1~9999.9 ml |
Total Volume | 0.0~9999.9 ml |
Time | 1min~99h00min |
Purge Rate | 600ml/h,display purge volume |
KVO | 1.0ml/h~5.0ml/h,adjustable |
Keypad | Numeric keypad |
BOLUS | Bolus function available, flflow rate adjustable |
Night Mode | Screen to be dim automatically in 5mins after standby time expired. |
Occlusion pressure detector | Occlusion pressure detection, 40KPa~130KPa, 10 levels adjustable with dynamic pressure value display |
Occlusion pressure release function | Available |
Air Bubble Detector | Ultrasonic detection, ≥50ul,Single bubble volume:50~300ul, 10 levels adjustable. |
Volume | 8 levels adjustable |
Motor Anti-reverse Function | Anti-reverse function and real-time monitoring |
IV Set Management | Open system, accept different lV brands aftercalibrationcalibration |
Self-Testing | Self-Testing before startup and during infusion |
Audio Visual Alarm Message | Air bubble, occlusion, dooropen, infusionfinish, standby time expired, battery low,battery empty, battery disengaged, AC Powerdisengaged, motor error, pump error, air sensorerror, parameter error, system error , lV notcalibrated |
Screen | 3.2inch brightness LCD |
Power Supply | AC Power:100V~240V,50Hz/60Hz,≤25VA |
Battery | Rechargeable Lithium battery, DC11.1V, 3000mAh;operating time: ≥10h @25ml/h,displaying remaining battery |
Operation Condition | Temperature:+5℃~+40℃, Humidity:20%~ 90%,Air pressure:70~106KPa |
Transport & Storage | Temperature :-20℃~+55℃,Humidity: 10%~95%,Air pressure:70~106KPa |
Classifification | Type CF, Class I, IPX3 |
Dimension | 145mm×150mm×200mm;Weight:1.7kg |